国产日韩欧美一区二区三区三州_亚洲少妇熟女av_久久久久亚洲av国产精品_波多野结衣网站一区二区_亚洲欧美色片在线91_国产亚洲精品精品国产优播av_日本一区二区三区波多野结衣 _久久国产av不卡

?

用于單分子動力學實驗的微流控混合器

2011-12-11 09:08:08支澤勇劉鵬程黃巖誼趙新生
物理化學學報 2011年8期
關鍵詞:核酸酶混合器微流

支澤勇 劉鵬程 黃巖誼 趙新生,*

(1北京大學化學與分子工程學院化學生物學系,北京分子科學國家實驗室,分子動態(tài)與穩(wěn)態(tài)結構國家重點實驗室,北京100871;2北京大學工學院,北京100871;3北京大學生物動態(tài)光學成像中心,北京100871)

用于單分子動力學實驗的微流控混合器

支澤勇1,3劉鵬程1,3黃巖誼2,3趙新生1,3,*

(1北京大學化學與分子工程學院化學生物學系,北京分子科學國家實驗室,分子動態(tài)與穩(wěn)態(tài)結構國家重點實驗室,北京100871;2北京大學工學院,北京100871;3北京大學生物動態(tài)光學成像中心,北京100871)

設計制作了用于單分子動力學實驗的微流控混合器,該混合器用聚二甲基硅氧烷(PDMS)芯片和石英載玻片密封而成,具有低的熒光背景,廣泛的生物相容性,結合激光共聚焦顯微鏡能夠在非平衡態(tài)下進行單分子熒光探測.我們設計的壓力控制系統(tǒng)和進樣流路方便而穩(wěn)定,保證了微流路中流形的長時間穩(wěn)定,從而實現(xiàn)了樣品流速和流量的精準控制.這些技術特點保證了單分子探測得到準確和高信噪比的結果.利用蛋白質的塌縮過程遠快于混合過程的特點,采用熒光標記的金黃色葡萄球菌核酸酶作為指示物,分辨出蛋白質變性態(tài)的特征峰,并利用變性態(tài)的熒光共振能量傳遞效率隨時間的變化表征出混合器在適合于單分子探測條件下的混合時間為150 ms.

微流控混合;單分子探測;熒光共振能量傳遞;蛋白質折疊;金黃色葡萄球菌核酸酶

1 Introduction

The protein folding study involves structure,thermodynamics,and kinetics.The kinetics of protein folding concerns the folding pathway,the rate,and the energy landscape.1-3In order to characterize the protein folding,experiments needs to be performed both under equilibrium and nonequilibrium conditions.4,5Single-molecule fluorescence resonance energy transfer (smFRET)can separate the subpopulations of the protein molecules beyond the capacity of common ensemble experiments.6-8Equilibrium smFRET experiments have addressed a number of important issues in protein folding by resolving the thermodynamic states and the sizes of the protein molecules within a heterogeneous mixture.9-12

Microfluidic laminar-flow mixers have been applied in many fields for kinetics measurements of biomolecular conformational changes with ultrafast mixing times.13-18Microfabricated mixers often utilize hydrodynamic focusing to squeeze the sample stream into submicron width to achieve extremely fast mixing through molecular diffusion.19Compared with traditional stopped flow method,the microfluidic mixer has advantages of submillisecond mixing time,greater uniformity,and low sample consumption.The mixing devices have the capacity of single-molecule fluorescence detection with the accessible window for high numerical aperture objectives.20-22Ensemble nonequilibrium experiments can only measure the kinetics of protein folding with the averaged and overall information, while the individual processes were indirectly resolved with kinetic modeling.23Single-molecule fluorescence detection under nonequilibrium conditions can be used to study the protein folding kinetics with the full distribution of conformations to separate the unfolded and folded states.24The combination of smFRET and a microfluidic mixer will generate novel insights into protein folding mechanism and is a powerful method to study biomolecular interactions and reactions.

Here,we constructed a microfluidic mixing system suitable for single-molecule fluorescence detection,which requires high signal-to-noise ratio,low protein adhesion,and stable flow rate for a long time.The mixing device was made of a cast of poly(dimethylsiloxane)(PDMS)sealed by a microscopic coverglass.The mixer channel pattern was designed to achieve comprehensive mixing at a minimal dead time and the flow was then slowed down to provide sufficient dwell time for single-molecule detection,and the device possessed wide biological compatibility.The mixer was easy to fabricate with common apparatus and could be improved for faster mixing with finer design and fabrication.The device was optimized to reduce the protein adhesion to the channel walls by using a long period of cure time and by adding wild-type protein at micromolar concentration into the injected sample.Novel design was implemented on the pressure regulation and sample-inlet lines so as to achieve a stable hydrodynamic flow in the mixer for many hours,which was a key factor for the single-molecule fluorescence measurements.To our knowledge,for the first time the mixing process along the mixing channel using smFRET histograms and the accurate flow velocity profile using fluorescence correlation spectroscopy(FCS)were characterized simultaneously.Then,the mixing time was determined by monitoring the collapse of denatured staphylococcal nuclease(SNase).25Although the mixing system was designed for single-molecule experiments,the mixer can also be used in ensemble measurements with a submillisecond time resolution by applying higher pressures.

2 Materials and methods

2.1 Construction of microfluidic mixing device

The channel drawing was created using computer-aided software.The patten was written by an electron beam on a chrome coated glass plate to generate the mask.The microfluidic mixer was made of a PDMS chip(RTV 615,GE Silicones)sealed to a No.1 coverglass(Fisher Scientific).The master was fabricated by contact photolithography.A 20 μm layer of SU-8 2010 (MicroChem,U.S.)was spin-coated onto a cleaned silicon wafer,after a soft bake the photoresist was exposed to UV light with 144 mJ·cm-2through the mask and the wafer was then baked on a hot-plate for 4.5 min at 95°C and developed.

In order to prevent PDMS adhesion,the master was silanized by exposure to a vapor of chlorotrimethylsilane (TMCS)in a sealed box for 15 min.To make the PDMS chip, a~4 mm thick layer of PDMS mixture,five parts by weight of PDMS and one part of crosslinking agent,was poured onto the mold and cured at 80°C for 10 min.The PDMS chip was then peeled off from the master,trimmed to the individual chip size and cured at 80°C for another 8 h.Finally,the PDMS chips were punched to generate inlet and outlet ports and bonded to a coverglass permanently using an air plasma.

2.2 Sample delivery system

A pressure-driven pump was used because it is applicable in the situation of low flow rate.26The pressure-driven sample delivery creates stable flow rate about 0.1 nL·s-1in the centre inlet for many hours.The protein sample and buffer were delivered into the inlets from two reservoirs which were made of 0.6 mL centrifuge tubes and PDMS plugs with two punched holes for Gauge 18 needles.The two sample reservoirs were connected to compressed air,and the pressures were regulated by two accurate pressure regulators(8286,Porter Instruments, U.S.)and measured with two accurate digital pressure gauges (DPG4000,Omega,U.S.)respectively.Another high pressure regulator having a range of 0-250 kPa was built to drive the solutions into the mixer in a faster flow speed,which can also be used in rapid mixing ensemble experiments.

In order to maintain the flow speed around 1 mm·s-1in the measurement channel for single-molecule detection,the side inlet pressure was regulated typically to 6.00 kPa and the centre inlet pressure was 7.80 kPa.The resolution of the regulators was about 0.01 kPa with a careful tuning.The pressures would change a little for individual mixers due to small deviation in the fabrication.The high pressure regulator was used to drive the solutions into the mixer at 150 kPa to focus the sample stream in a short time.If the solutions were pumped into the mixer by the precise but low pressure regulators,hours would be needed to obtain a stable focused sample stream.After the formation of the focused stream,the pressures of the reservoirs were switched to the precise and low pressure regulators for single-molecule experiments.

2.3 Single-molecule confocal microscope

Single-molecule fluorescence measurements were performed on a home-built dual-channel confocal fluorescence microscope27,28based on a TE2000 microscope(Nikon).The labeled protein sample was excited by a solid-state laser(MLLIII-532,CNI)at 532 nm with 100 μW for experiments on the coverglass and 130 μW in the mixer,focused through an oil immersion objective(100×,NA 1.3,Nikon,Japan).The donor and acceptor fluorescences,seperated from the excitation light by a dichroic mirror(Z532,Chroma,U.S.),were collected by the same objective and spatially filtered using a 30 μm pinhole. The passed fluorescence was separated into donor and acceptor components with a second dichroic mirror(FF650-Di01,Semrock,U.S.)and two final filters(FF01-593/40 and FF01-692/ 40,Semrock,U.S.)for the donor and acceptor channels,respectively.Each component was detected by a photon-counting Avalanche Photodiode(SPCM-AQRH-14,PerkinElmer Optoelectronics,U.S.).Fluorescence intensities were recorded with a photon counters card(PMS-400A,Becker&Hickl,Germany). Autocorrelation functions were simultaneously recorded using a multiple-digital hardware correlator device(Flex02-01D, www.correlator.com,U.S.).

The raw single-molecule fluorescence data were corrected for several factors11,29to obtain the actual FRET efficiency,including background,differences in quantum yields,different collection efficiencies of the donor and acceptor channels, cross-talk,and direct excitation of the acceptor.

2.4 Protein expression,purification and labeling

Expression,purification,and labeling of the mutant of SNase,K28C-H124C,were carried out as described previously.30Briefly,The mutant proteins were reduced with excess of Dithiothreitol(DTT,Sigma,U.S.)followed by chromatography in labeling buffer to remove the excess DTT.Site-specific labeling was achieved by reaction with thiol-reactive fluorescence dyes Alexa Fluor 555 and Alexa Fluor 647(Invitrogen, U.S.).Free dyes were removed through a PD-10 Desalting Column(GE Healthcare,U.S.)and the labeled protein solution was stored at-80°C with 10%glycerol.

Labeled SNase of 50-100 pmol·L-1was diluted in 1 μmol·L-1unlabeled SNase in Tris-HCl buffer(pH 7.8,50 mmol·L-1Tris-HCl with 100 mmol·L-1NaCl)at appropriate GdmCl concentration.The dwell time bin was 1 ms10and a threshold was set at 50 counts in the sum of photon counts from the two channels.

3 Results and discussion

3.1 Controbility of the mixer

The microfluidic mixer shown in Fig.1 was designed with a resistance model using Ohm?s law.19The relationship of flow rate,pressure difference,and flow impedance in a rectangular pipe can be described with the following equation:31

where Q is the flow rate,dp/dl is the pressure gradient,w is the channel width,h is the channel height,and η is the solution viscosity.The two side channels of an actual microfluidic mixer were connected to the same entry,reducing a set of pressure regulator system.The channel dimensions and the flow impedances are shown in Table 1.The impedance of the mixing neck is less than 1%of the other channels which can be ignored in the calculation.When the change of inlets pressure is 0.05 kPa, the maximum pressure fluctuation of the system,the change in flow ratio is about 2%,so that the mixer can maintain a stable flow rate for smFRET measurement.

Fig.1 (A)Scheme of the microfluidic mixer,(B)a white light microscopy image of the microfluidic mixing region,(C)the impedance model for the designed mixerThe height of the channels in figure B is 20 μm.In figure C:Zcand Zmare the impedances of the centre and measurement channels,respectively;Zsis the impedance equivalent to the side flow impedance of a mixer with two parallel side channels.

Table 1 Channel dimensions and impedances

Fig.2 Flow stability in the microfluidic mixer(A)fluorescence images of focused dye solution stream over hours;The intensity profiles of the cross line(the white line)for all images are extracted. (B)The peak intensities of the cross lines are plotted as time,which is 11270±280,indicating that the flow rate fluctuation was about 2%.

The robustness of the flow velocity was examined by fluorescence images over several hours.The pressure was set to maintain appropriate hydrodynamic focusing and a flow rate in the detection channel around 1 mm·s-1.A 0.1 μmol·L-1AF532 solution with 0.01%Tween 20(Pierce)to prevent dye adhesion was delivered into the centre inlet and water was injected into the side inlets.The focused dye solution stream was excited by mercury lamp and recorded by a CCD camera through an emission filter(Fig.2).The high stability of the flow rate guaranteed the fidelity of single-molecule measurements over a long time for a good signal-to-noise ratio.When a denatrued potein sample with 2 mol·L-1GdmCl was fed into the centre inlet,simulation reslults showed that the concentration deviation of the denatrurant was less than 0.05 mol·L-1.As a consequence,the protein concentration and conformation were stable at the focus of the optics.

3.2 Flow velocity profile

In order to convert the focus position into the corresponding time,the flow velocity in the channels was measured by FCS.32,33Advantage of this approach is the large measurable velocity range,from 0.1 mm·s-1to 10 m·s-1,and the simultaneous measurement of the flow rate with the collection of smFRET events.FCS measured the fluorescent molecule number fluctuations in a small focal volume15about 1 fL.In the mixing experiments,FCS curves(Fig.3A)were fit using a model of considering diffusion,flow,and singlet-triplet transition of the fluorescent molecule,27,33

where N is the average number of molecules in the focus volume,τdiffis the characteristic diffusion time,τ/τflowis the characteristic flow time,K is the fraction of the triplet state,and τTis the characteristic triplet state time.τflowis obtained from the fitting for each FCS curve,which is related to the flow speed by

where r0is the radius of the focus volume,which is determined using Rhodamine 6G34to be(270±7)nm.

We measured the velocities along the central axis of the channels.In an smFRET experiment each fluorescent burst needs at least dozens of effective photons35and requires the flow rate to be about 1 mm·s-1so that each molecule can spend about 1 ms at the focus.Fig.3B shows the velocity distri-bution along the central axis adjusted at such a condition.

Fig.3 (A)Autocorrelation curves at different positions in the device,(B)the velocity profile in the mixerIn figureA:the beginning point of the measurement channel is 0 μm(Fig.1B).-165 μm is located in the centre channel with the minimum flow rate, -50 μm is in the mixing neck with the maximum flow rate to achieve rapid mixing,and 500 μm is in the measurement channel with a flow rate of about 1 mm·s-1,suitable for smFRET experiments.

Fig.4 Single-molecule fluorescence traces(A)single-molecule signal of a dye solution of 10 pmol·L-1AF546 on a coverglass and in the mixer,(B)raw smFRET data of 100 pmol·L-1AF555 and AF647 dual labeled SNase in the mixer channel,excited by a 532 nm laser

3.3 Single-molecule detection

To test the fluorescence collection efficiency and the background level in the mixer channel,the single-molecule fluorescence trace of a 10 pmol·L-1dye solution of Alexa Fluor(AF) 546 in water was collected both on a coverglass and in the mixer channel in static experiments(Fig.4A).The dye solution was excited by a 532 nm laser at a power of 100 μW,and the collected raw data were binned into 1 ms dwell time.The background was fitted by a Poisson distribution,and the average photon counting rates are 0.32 m·s-1on a coverglass and 0.81 m·s-1in the mixer channel.

We also tested the data quality for a protein sample labeled with a dye pair in the mixer channel.A 100 pmol·L-1AF555 and AF647(Invitrogen)dual labeled SNase30mixed with 1 μmol·L-1wild-type SNase to prevent labeled protein adhesion was delivered into the microfluidic device.The sample was excited with a 130 μW laser beam to generate FRET events (Fig.4B).The background levels for the donor channel and acceptor channel were 1.5 and 1.2 m·s-1,respectively,which were higher than the pure dye solution due to impurities and higher concentration,but the signal levels were higher as well, resulting in even better signal-to-noise ratio of nearly 100 for the protein sample.

3.4 Mixing time characterized using collapse of denatured SNase

When unfolded protein is transferred from high denaturant concentration to low concentration,the size of the protein molecule will collapse which leads to higher intramolecular smFRET efficiency and the protein collapse time is about several hundred nanoseconds,36much faster than the mixing time. This property was used to measure the mixing time precisely. Fig.5A shows the smFRET histograms taken at different positions in the channels,fitted by lognormal and Gaussian distributions to obtain the apparent FRET efficiency of the unfolded state.The rectangular box in Fig.5B indicates the onset and completion of the mixing.The initial position was set to Eapp= 0.42,5%larger than the Eapp=0.40 before mixing,and the end point was set to the first accessible position for single-molecule detection.So doing,we found that the mixing time was 150 ms.The simplest way to enhance the mixing process is to minimize the width of mixing regions with better manufacture. Because the adequate dwell time for single-molecule detection is about 1 ms for fluorescent molecules flowing in the focus volume,the ultrafast mixing can be realized by a design of flow velocity deceleration.22The mixercombining with smFRET for kinetics study has been applied to the measurement of the folding rate of denatured SNase.We will report our results in a future publication.

Fig.5 Mixing process elucidated by smFRET experiments(A)smFRET histograms along the central axis of the channels,(B)the fitted FRET efficiencies of the denatured SNase; The rectangular box indicates the mixing region with a mixing time tmixof 150 ms.

4 Conclusions

A microfluidic mixing system was constructed especially for single-molecule kinetic measurement,which has the advantages of high signal-to-noise ratio,stable flow rate over hours,accurate time determination by employing FCS,and easy fabrication.This technique has the capacity to study a wide variety of biological reactions requiring the combination of fast mixing, single-molecule detection,and small sample consumption.The construction of the mixer makes it possible for us to study the kinetics of various biochemical processes at a single-molecule level.

(1) Wolynes,P.G.;Onuchic,J.N.;Thirumalai,D.Science 1995, 267,1619.

(2) Oliveberg,M.;Wolynes,P.G.Q.Rev.Biophys.2005,38,245.

(3) Ferreon,A.C.M.;Deniz,A.A.BBA-Proteins Proteomics 2011, in press.

(4)Haas,E.ChemPhysChem 2005,6,858.

(5) Bilsel,O.;Matthews,C.R.Curr.Opin.Struct.Biol.2006,16, 86.

(6) Ha,T.;Enderle,T.;Ogletree,D.F.;Chemla,D.S.;Selvin,P.R.; Weiss,S.Proc.Natl.Acad.Sci.U.S.A.1996,93,6264.

(7) Weiss,S.Science 1999,283,1676.

(8)Deniz,A.A.;Mukhopadhyay,S.;Lemke,E.A.J.R.Soc. Interface 2008,5,15.

(9) Deniz,A.A.;Laurence,T.A.;Beligere,G.S.;Dahan,M.; Martin,A.B.;Chemla,D.S.;Dawson,P.E.;Schultz,P.G.; Weiss,S.Proc.Natl.Acad.Sci.U.S.A.2000,97,5179.

(10) Schuler,B.;Lipman,E.A.;Eaton,W.A.Nature 2002,419,743.

(11) Hoffmann,A.;Kane,A.;Nettels,D.;Hertzog,D.E.; Baumgartel,P.;Lengefeld,J.;Reichardt,G.;Horsley,D.A.; Seckler,R.;Bakajin,O.;Schuler,B.Proc.Natl.Acad.Sci.U.S. A.2007,104,105.

(12) Muller-Spath,S.;Soranno,A.;Hirschfeld,V.;Hofmann,H.; Ruegger,S.;Reymond,L.;Nettels,D.;Schuler,B.Proc.Natl. Acad.Sci.U.S.A.2010,107,14609.

(13) Hertzog,D.E.;Michalet,X.;Jager,M.;Kong,X.X.;Santiago, J.G.;Weiss,S.;Bakajin,O.Anal.Chem.2004,76,7169.

(14)Hertzog,D.E.;Ivorra,B.;Mohammadi,B.;Bakajin,O.; Santiago,J.G.Anal.Chem.2006,78,4299.

(15) Park,H.Y.;Qiu,X.Y.;Rhoades,E.;Korlach,J.;Kwok,L.W.; Zipfel,W.R.;Webb,W.W.;Pollack,L.Anal.Chem.2006,78, 4465.

(16) Lapidus,L.J.;Yao,S.H.;McGarrity,K.S.;Hertzog,D.E.; Tubman,E.;Bakajin,O.Biophys.J.2007,93,218.

(17) Park,H.Y.;Kim,S.A.;Korlach,J.;Rhoades,E.;Kwok,L.W.; Zipfell,W.R.;Waxham,M.N.;Webb,W.W.;Pollack,L.Proc. Natl.Acad.Sci.U.S.A.2008,105,542.

(18)Guo,S.;Xue,M.Q.;Qian,M.X.;Cao,T.B.;Zhao,X.S.Acta Phys.-Chim.Sin.2007,23,1827.[郭 素,薛面起,錢民協(xié),曹廷炳,趙新生.物理化學學報,2007,23,1827.]

(19) Knight,J.B.;Vishwanath,A.;Brody,J.P.;Austin,R.H.Phys. Rev.Lett.1998,80,3863.

(20) Hamadani,K.M.;Weiss,S.Biophys.J.2008,95,352.

(21)Pfeil,S.H.;Wickersham,C.E.;Hoffmann,A.;Lipman,E.A. Rev.Sci.Instrum.2009,80,055105.

(22)Gambin,Y.;VanDelinder,V.;Ferreon,A.C.M.;Lemke,E.A.; Groisman,A.;Deniz,A.A.Nat.Methods 2011,8,239.

(23) Maki,K.;Cheng,H.;Dolgikh,D.A.;Roder,H.J.Mol.Biol. 2007,368,244.

(24) Lipman,E.A.;Schuler,B.;Bakajin,O.;Eaton,W.A.Science 2003,301,1233.

(25) Ye,K.Q.;Wang,J.F.J.Mol.Biol.2001,307,309.

(26) Kim,S.J.;Blainey,P.C.;Schroeder,C.M.;Xie,X.S.Nat. Methods 2007,4,397.

(27) Krichevsky,O.;Bonnet,G.Rep.Prog.Phys.2002,65,251.

(28)Chen,X.D.;Zhou,Y.;Qu,P.;Zhao,X.S.J.Am.Chem.Soc. 2008,130,16947.

(29) Sherman,E.;Haran,G.Proc.Natl.Acad.Sci.U.S.A.2006, 103,11539.

(30) Liu,P.C.;Meng,X.L.;Qu,P.;Zhao,X.S.;Wang,C.C. J.Phys.Chem.B 2009,113,12030.

(31) White,F.Viscous Fluid Flow,2nd ed.;McGraw Hill:Boston, Massachusetts,1991.

(32) Gosch,M.;Blom,H.;Holm,J.;Heino,T.;Rigler,R.Anal. Chem.2000,72,3260.

(33) Kuricheti,K.K.;Buschmann,V.;Weston,K.D.Appl. Spectrosc.2004,58,1180.

(34)Nie,S.M.;Chiu,D.T.;Zare,R.N.Anal.Chem.1995,67,2849.

(35) Gell,C.;Brockwell,D.;Smith,A.Handbook of Single Molecule Fluorescence Spectroscopy;Oxford University:Oxford,2006.

(36) Nettels,D.;Gopich,I.V.;Hoffmann,A.;Schuler,B.Proc.Natl. Acad.Sci.U.S.A.2007,104,2655.

April 27,2011;Revised:June 1,2011;Published on Web:June 9,2011.

A Microfluidic Mixer for Single-Molecule Kinetics Experiments

ZHI Ze-Yong1,3LIU Peng-Cheng1,3HUANG Yan-Yi2,3ZHAO Xin-Sheng1,3,*
(1Beijing National Laboratory for Molecular Sciences,State Key Laboratory for Structural Chemistry of Unstable and Stable Species,Department of Chemical Biology,College of Chemistry and Molecular Engineering,Peking University,Beijing 100871, P.R.China;2College of Engineering,Peking University,Beijing 100871,P.R.China;3Biodynamic Optical Imaging Center, Peking University,Beijing 100871,P.R.China)

We designed and built a microfluidic mixer based on the principle of hydrodynamic focusing governed by Navier-Stokes equation for single-molecule kinetics experiments.The mixer is a cast of poly(dimethylsiloxane)(PDMS)sealed with transparent fused-silica coverglass,which results in low fluorescence background and broad biological compatibility and this enables single-molecule fluorescence detection under nonequilibrium conditions.The pressure regulated sample delivery system is convenient for loading a sample and allows for precise and stable flow velocity control.The combination of microfluidic mixer and single-molecule fluorescence resonance energy transfer(smFRET)allows us to measure the time course of the distribution of the smFRET efficiency in protein folding.We used the fact that denatured protein collapses much faster than the mixing process to characterize the mixing time using donor and acceptor dyes labeled staphylococcal nuclease(SNase)as an smFRET efficiency indicator.By monitoring the smFRET efficiency of denatured SNase during the course of mixing,we determined that the mixing time was 150 ms under conditions suitable for single-molecule detection.

Microfluidic mixing;Single-molecule detection;Fluorescence resonance energy transfer; Protein folding;Staphylococcal nuclease

?Corresponding author.Email:zhaoxs@pku.edu.cn;Tel:+86-10-62751727.

The project was supported by the National Natural Science Foundation of China(20733001,20973015)and National Key Basic Research Program of China(973)(2006CB910300,2010CB912302).

國家自然科學基金(20733001,20973015)和國家重點基礎研究發(fā)展規(guī)劃項目(973)(2006CB910300,2010CB912302)資助

O643

猜你喜歡
核酸酶混合器微流
粘質沙雷氏菌全能核酸酶的研究進展
船用發(fā)動機SCR混合器優(yōu)化仿真分析
含季銨鹽的芳酰腙配體的銅 (Ⅱ)配合物的合成和表征:體外DNA鍵合和核酸酶活性
多種Cas12a蛋白變體能識別不同的PAM序列(2020.4.27 Plant Biotechnology Journal)
微流控法制備P(NIPA-co-MAA)水凝膠微球及其性能表征
用megaTAL 核酸酶對原代人T 細胞CCR5 基因座進行有效修飾可建立HIV-1 抵抗力
旋轉盤式混合器混合過程數值模擬
中國塑料(2016年7期)2016-04-16 05:25:54
微流控芯片在食品安全分析中的應用進展
微流控SERS芯片的設計制備及其在細菌檢測中的應用
紙芯片微流控技術的發(fā)展及應用
抚松县| 陇川县| 潞城市| 墨脱县| 南靖县| 左贡县| 万安县| 霍林郭勒市| 万全县| 资溪县| 阳山县| 曲松县| 南皮县| 广德县| 长兴县| 永安市| 阿拉善右旗| 东乌珠穆沁旗| 响水县| 霍林郭勒市| 白城市| 安平县| 台安县| 新平| 仙桃市| 白沙| 深圳市| 进贤县| 玛沁县| 福建省| 韶关市| 宜兰县| 桐柏县| 温宿县| 沛县| 文昌市| 红河县| 新龙县| 万年县| 岱山县| 肥乡县|