祝安惠,張燕燕
(北京大學(xué)第三醫(yī)院核醫(yī)學(xué)科,北京 100191)
PET/MR是將PET和MR融為一體的影像學(xué)檢查手段。PET/MR集PET功能成像和MR精準(zhǔn)定位及多參數(shù)成像等優(yōu)勢于一體,隨著圖像重建、衰減矯正及偽影控制技術(shù)的提高,其掃描時間逐步縮短,輻射劑量進(jìn)一步降低,在臨床及科研中的應(yīng)用領(lǐng)域逐年拓寬[1]。本文對一體化PET/MR的應(yīng)用進(jìn)展進(jìn)行綜述。
隨著影像融合技術(shù)的不斷進(jìn)步,PET/MR成像設(shè)備已由傳統(tǒng)的分體式發(fā)展成為一體機(jī),其構(gòu)造及性能顯著提升,克服了分體式PET/MR檢查過程耗時長及無法實現(xiàn)同時掃描的不足。一體化PET/MR系統(tǒng)設(shè)計分為插入式、嵌入式及完全集成式3種方式[2]。①插入式:將PET探測器簡單插入MR設(shè)備,但PET與MR信號相互干擾程度較大,導(dǎo)致PET性能下降;②嵌入式:將PET探測器置于MR發(fā)射線圈內(nèi),可實現(xiàn)同步掃描,但也導(dǎo)致MR孔徑明顯變小而僅可用于小動物或人體頭部和四肢;③完全集成式:將PET探測器置入MR梯度線圈內(nèi),可節(jié)省磁體占據(jù)的空間,同時顯著降低PET與MR互相干擾,使PET性能顯著提高,可實現(xiàn)飛行時間(time of flight, TOF)技術(shù),真正實現(xiàn)PET與MR等中心、相同容積和同步掃描,使PET/MR整體掃描時間明顯縮短[3]。
1.1 硬件技術(shù) 影響一體化PET/MR性能的硬件技術(shù)主要為PET光電轉(zhuǎn)換器、MR線圈性能及連續(xù)床位移動技術(shù)等。PET光電轉(zhuǎn)換器有光電倍增管(photomultiplier tubes, PMT)、雪崩光電二極管(avalanche photo diode, APD)和硅光電倍增管(silicon photomultipliers, SiPM)。PMT對磁場敏感而不適用于PET/MR設(shè)備;APD對磁場不敏感,但噪聲較大而導(dǎo)致能量和時間分辨率下降,在一體化PET/MR設(shè)備中已逐漸被SiPM取代。SiPM的光電轉(zhuǎn)化率、時間和空間分辨率均較APD高,且工作電壓低而增益高,可實現(xiàn)高度集成化,進(jìn)一步提高PET的整體性能[4]。目前基于SiPM的TOF時間分辨率已提高至400 ps以內(nèi)[5]。MR采用一體化相控陣表面線圈,配合連續(xù)床位移動技術(shù),一次采集可同時獲得全身高分辨率圖像及T1W、T2W、質(zhì)子加權(quán)和DWI等多序列圖像,掃描速度明顯提高[1]。
1.2 軟件技術(shù) 為提高PET圖像信噪比和對小病灶的檢出率,一體化PET/MR軟件技術(shù)的突破點主要在于圖像重建和衰減校正方法。PET采用點擴(kuò)展函數(shù)(point spread function, PSF)重建方法,可獲得亞毫米分辨率。有學(xué)者[6]發(fā)現(xiàn),對于最大徑<10 mm的病灶,TOF聯(lián)合PSF重建方法獲得的標(biāo)準(zhǔn)攝取平均值和最大值均較單獨使用TOF或PSF提高,提示基于TOF聯(lián)合PSF重建方法的一體化PET/MR有利于檢出最大徑<10 mm的病灶。TOF技術(shù)可提高掃描速度及PET圖像對比度,從而降低示蹤劑用量[4]、提高圖像質(zhì)量[7]。
一體化PET/MR采用MR信息對PET圖像進(jìn)行衰減校正(MR attenuation correction, MRAC)[8],包括提高組織分類、恢復(fù)人體輪廓和提高衰減校正圖的準(zhǔn)確性。準(zhǔn)確地組織分類是精確計算衰減校正系數(shù)圖的前提,目前MR通過水脂分離序列、圖集技術(shù)及零回波技術(shù)可獲得5種組織分類(氣體、水、脂肪、軟組織和骨骼),聯(lián)合應(yīng)用上述技術(shù)可顯著提高M(jìn)RAC的可靠性、可重復(fù)性和準(zhǔn)確性[9]。
1.3 偽影校正 一體化PET/MR偽影校正主要針對運動偽影及金屬植入物相關(guān)偽影。PET和MR運動偽影校正技術(shù)可啟動PET與MR同步掃描,包括外置門控(如呼吸、心電和指脈信號等)和MR自身導(dǎo)航技術(shù)。一體化TOF PET/MR不增加掃描時間即可提供與呼吸周期不同時相匹配的MRAC校正技術(shù),可明顯減少由呼吸運動導(dǎo)致的圖像不匹配偽影,從而顯著提高胸腹部圖像質(zhì)量[10]。TOF技術(shù)還可補(bǔ)償金屬植入物所致的PET信息丟失,提高M(jìn)RAC的精確性[11]。
2.1 腫瘤
2.1.1 肺癌 一體化TOF PET/MR的亞毫米級時間分辨力能有效消除運動偽影、顯著提高圖像分辨力。研究[12]表明,一體化TOF PET/MR能有效檢出肺部病變,可應(yīng)用于肺癌的臨床分期。
2.1.2 乳腺癌 乳腺鉬靶和超聲難以判斷致密腺體中可能存在的腫塊。MR具有優(yōu)越的軟組織分辨力,可準(zhǔn)確檢出此類病變,且在顯示乳腺癌胸壁侵犯方面具有明顯優(yōu)勢。有研究[13-14]報道,俯臥位PET/MR乳腺成像中,雙乳處于下垂?fàn)顟B(tài),所提供的病變代謝、解剖細(xì)節(jié)及功能參數(shù)方面的信息均優(yōu)于常規(guī)仰臥位PET/MR成像。
2.1.3 腹部及盆腔腫瘤 PET/MR有利于檢出高分化肝癌、胃腸道神經(jīng)內(nèi)分泌腫瘤、女性生殖系統(tǒng)原發(fā)腫瘤、轉(zhuǎn)移病變及腹腔小淋巴結(jié)[15],還可準(zhǔn)確評估胰腺癌的侵犯范圍。有學(xué)者[1]采用11C或18F標(biāo)記膽堿或68Ga-PSMA,發(fā)現(xiàn)PET/MR能清晰顯示前列腺癌原發(fā)病灶的解剖細(xì)節(jié),有助于指導(dǎo)穿刺活檢。
2.1.4 神經(jīng)系統(tǒng)腫瘤 PET/MR可獲得病變的精細(xì)解剖細(xì)節(jié)及代謝信息,還可通過MR擴(kuò)散張量成像、DWI、MRS等多參數(shù)成像獲得腫瘤侵犯神經(jīng)纖維束、腫瘤內(nèi)部微環(huán)境變化及腫瘤化學(xué)成分變化的相關(guān)信息[16],在手術(shù)或活檢前精確定位腫瘤方面具有重要意義。
2.1.5 淋巴瘤 MR具有優(yōu)越的軟組織分辨率,不僅可發(fā)現(xiàn)局限在淋巴結(jié)的病變,還可有效檢出骨髓、肌肉、腹部及盆腔臟器等的微小浸潤灶。由于MR無輻射,適宜放化療后再分期、療效評估和遠(yuǎn)期隨訪,尤其是對射線敏感的兒童患者,更是理想的選擇[17]。
2.1.6 骨轉(zhuǎn)移瘤 MR快速自旋回波T1WI或Dixon同相位T1WI對骨髓浸潤病變非常敏感,有助于檢出早期骨轉(zhuǎn)移瘤,同時可用于評估骨轉(zhuǎn)移瘤放化療療效[18]。
2.2 非腫瘤病變
2.2.1 神經(jīng)系統(tǒng) 一體化TOF PET/MR可提供腦組織精細(xì)化的解剖結(jié)構(gòu)、腦血流量及腦功能等多方面的信息,實現(xiàn)PET絕對定量腦血流量,在早期診斷腦疾病(如腦血管病、癡呆、帕金森病及癲癇等)導(dǎo)致的腦血流量異常及預(yù)后判斷中發(fā)揮重要作用[19]。Su等[20]發(fā)現(xiàn),采用15O-PET圖像衍生的動脈輸入技術(shù),PET/MR可絕對定量腦血流動力學(xué)參數(shù)(如腦血流量、血容量及腦氧代謝率),且準(zhǔn)確性與傳統(tǒng)侵入性動脈血采樣法相同,操作簡單,可重復(fù)性強(qiáng)。一體化PET/MR能識別腦缺血時發(fā)生無氧糖酵解的組織,可用于觀察缺血后神經(jīng)炎癥導(dǎo)致神經(jīng)破壞和修復(fù)發(fā)生的機(jī)制。Lewis等[21]采用一體化52Mn PET/MR檢查小鼠神經(jīng)干細(xì)胞,為未來臨床研究神經(jīng)干細(xì)胞治療技術(shù)及治療策略提供影像學(xué)依據(jù)。
2.2.2 心血管系統(tǒng) PET/MR輻射劑量低,具備良好的運動偽影矯正方法,同時心臟MR可提供精細(xì)的形態(tài)學(xué)信息,因此PET/MR在評價冠狀動脈不穩(wěn)定斑塊、心功能、心肌灌注、心肌存活及心肌病等方面具有重要作用,有望成為心血管疾病的常規(guī)檢查方法[22]。Robson等[23]采用自由呼吸衰減矯正成像方法和增加迭代次數(shù)的重建方法,應(yīng)用18F-FDG和18F-NaF兩種顯像劑,獲得冠狀動脈炎癥活動和微鈣化的低劑量(7.2 mSv)PET/MR圖像。新型示蹤劑18F-galacto-RGD可用于評估冠狀動脈斑塊炎癥及再生血管。
有學(xué)者[24]提出,MR不同序列檢測肺部病變的效能不同,故優(yōu)化MR采集序列可提高PET/MR的檢測效能。多項研究[25-26]顯示,ADC值和SUV值是診斷宮頸鱗癌的獨立指標(biāo),以ADC值評價惡性程度分級較SUV值更敏感,提示以DWI作為PET/MR常規(guī)掃描序列可為診斷及制定放療計劃提供參考信息。國內(nèi)已有學(xué)者制定了神經(jīng)系統(tǒng)PET/MR檢查規(guī)范[27]及護(hù)理流程[28],但國內(nèi)一體化PET/MR應(yīng)用尚處于起步階段,仍需進(jìn)一步規(guī)范掃描方案。
PET主要采用18F-FDG為探針,18F-FDG在多數(shù)惡性腫瘤中均表現(xiàn)為不同程度攝取增高。MR可通過多參數(shù)和多序列掃描來提高組織分辨率。采用18F-FDG可能導(dǎo)致漏檢18F-FDG攝取陰性或MRI常規(guī)成像缺乏良好對比的病變或組織,故研制既適用于MR又適用于PET的新型探針,將有助于進(jìn)一步拓寬PET/MR的應(yīng)用范圍。
[1] Spick C, Herrmann K, Czernin J.18F-FDG PET/CT and PET/MRI perform equally well in cancer: Evidence from studies on more than 2,300 patients. J Nucl Med, 2016,57(3):420-430.
[2] Delso G, Ziegler S. PET/MRI system design. Eur J Nucl Med Mol Imaging, 2009,36(Suppl 1):S86-S92.
[3] Levin CS, Maramraju SH, Khalighi MM, et al. Design features and mutual compatibility studies of the time-of-flight PET capable GE SIGNA PET/MR system. IEEE Trans Med Imaging, 2016,35(8):1907-1914.
[4] Oehmigen M, Ziegler S, Jakoby BW, et al. Radiotracer dose reduction in integrated PET/MR: Implications from National Electrical Manufacturers Association phantom studies. J Nucl Med, 2014,55(8):1361-1367.
[5] Grant AM, Deller TW, Khalighi MM, et al. NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system. Med Phys, 2016,43(5):2334.
[6] Shang K, Cui B, Ma J, et al. Clinical evaluation of whole-body oncologic PET with time-of-flight and point-spread function for the hybrid PET/MR system. Eur J Radiol, 2017,93(8):70-75.
[7] Kolb A, Sauter AW, Eriksson L, et al. Shine-through in PET/MR imaging: Effects of the magnetic field on positron range and subsequent image artifacts. J Nucl Med, 2015,56(6):951-954.
[8] Leynes AP, Yang J, Shanbhag DD. Hybrid ZTE/dixon MR-based attenuation correction for quantitative uptake estimation of pelvic lesions in PET/MRI. Med Phys, 2017,44(3):902-913.
[9] Sekine T, Burgos N, Warnock G, et al. Multi-atlas-based attenuation correction for brain18F-FDG PET imaging using a time-of-flight PET/MR scanner: Comparison with clinical single-atlas- and CT-based attenuation correction. J Nucl Med, 2016,57(8):1258-1264.
[10] Delso G, Khalighi M, Ter Voert E, et al. Effect of time-of-flight information on PET/MR reconstruction artifacts: Comparison of free-breathing versus breath-hold MR-based attenuation correction. Radiology, 2017,282(1):229-235.
[11] Defrise M, Rezaei A, Nuyts J. Time-of-flight PET data determine the attenuation sinogram up to a constant. Phys Med Biol, 2012,57(4):885-899.
[12] Minamimoto R, Iagaru A, Jamali M, et al. Conspicuity of malignant lesions on PET/CT and simultaneous time-of-flight PET/MRI. PLoS One, 2017,12(1):e0167262.
[13] Schaarschmidt BM, Grueneisen J, Metzenmacher M, et al. Thoracic staging with18F-FDG PET/MR in non-small cell lung cancer-does it change therapeutic decisions in comparison to18F-FDG PET/CT? Eur Radiol, 2017,27(2):681-688.
[14] Cho IH, Kong EJ. Potential clinical applications of18F-Fluorodeoxyglucose positron emission tomography/magnetic resonance mammography in breast cancer. Nucl Med Mol Imaging, 2017,51(3):217-226.
[15] Oprea-Lager DE, Yaqub M, Pieters IC, et al. A clinical and experimental comparison of time of flight PET/MRI and PET/CT systems. Mol Imaging Biol, 2015,17(5):714-725.
[16] Heiss WD. Hybrid PET/MR imaging in neurology: Present applications and prospects for the future. J Nucl Med, 2016,57(7):993-995.
[17] Afaq A, Fraioli F, Sidhu H, et al. Comparison of PET/MRI with PET/CT in the evaluation of disease status in lymphoma. Clin Nucl Med, 2017,42(1):e1-e7.
[18] Eiber M, Takei T, Souvatzoglou M, et al. Performance of whole-body integrated18F-FDG PET/MR in comparison to PET/CT for evaluation of malignant bone lesions. J Nucl Med, 2014,55(2):191-197.
[19] 單藝,盧潔,李坤成.一體化PET/MR評估腦血流量的研究進(jìn)展.中國醫(yī)學(xué)影像技術(shù),2017,33(8):1269-1272.
[20] Su Y, Vlassenko AG, Couture LE, et al. Quantitative hemodynamic PET imaging using image-derived arterial input function and a PET/MR hybrid scanner. J Cereb Blood Flow Metab, 2017,37(4):1435-1446.
[21] Lewis CM, Graves SA, Hernandez R, et al.52Mn production for PET/MRI tracking of human stem cells expressing divalent metal transporter 1 (DMT1). Theranostics, 2015,5(3):227-239.
[22] Kero T, Nordstr?m J, Harms HJ, et al. Quantitative myocardial blood flow imaging with integrated time-of-flight PET-MR. EJNMMI Phys, 2017,4(1):1.
[23] Robson PM, Dweck MR, Trivieri MG, et al. Coronary artery PET/MR imaging: Feasibility, limitations, and solutions. JACC Cardiovasc Imaging, 2017,10(10 Pt A):1103-1112.
[24] Rasmussen JH, N?rgaard M, Hansen AE, et al. Feasibility of multiparametric imaging with PET/MR in head and neck squamous cell carcinoma. J Nucl Med, 2017,58(1):69-74.
[25] 王鵬遠(yuǎn),辛軍,孫洪贊,等.18FDG-PET/MR觀察宮頸鱗狀細(xì)胞癌ADC值與SUV的相關(guān)性.中國醫(yī)學(xué)影像技術(shù),2014,30(4):603-607.
[26] Huellner MW, Appenzeller P, Kuhn FP, et al. Whole-body nonenhanced PET/MR versus PET/CT in the staging and restaging of cancers: Preliminary observations. Radiology, 2014,273(3):859-869.
[27] 盧潔,張苗,方繼良,等.一體化PET/MR顱腦成像檢查規(guī)范(2017版).中國醫(yī)學(xué)影像技術(shù),2017,33(5):791-794.
[28] 韓斌如,帥冬梅,方繼良,等.一體化PET/MR檢查臨床護(hù)理操作規(guī)范(2017).中國醫(yī)學(xué)影像技術(shù),2017,33(5):795-798.
《中華介入放射學(xué)電子雜志》贈刊、征稿啟事
《中華介入放射學(xué)電子雜志》(刊號:ISSN 2095-5782,CN 11-9339/R)創(chuàng)刊于2013年,由國家衛(wèi)生計生委主管、中華醫(yī)學(xué)會主辦的介入醫(yī)學(xué)專業(yè)學(xué)術(shù)期刊。是中華醫(yī)學(xué)會放射學(xué)分會介入學(xué)組指定的官方學(xué)術(shù)刊物??偩庉嫗閱硒櫧淌?,常務(wù)總編輯為翟仁友教授,以光盤附紙質(zhì)導(dǎo)讀形式公開發(fā)行出版。雜志為季刊,每期80頁左右,大16開本,銅版紙印刷,印刷精良,圖片清晰。
本刊主要刊登神經(jīng)介入、血管介入、腫瘤介入和非血管介入領(lǐng)域的臨床及科研新進(jìn)展,同時涵蓋介入護(hù)理、介入學(xué)科管理及教學(xué)、醫(yī)學(xué)影像等方面的研究成果。是廣大介入放射工作者展示學(xué)術(shù)成果、臨床經(jīng)驗和學(xué)習(xí)、交流的平臺。
本刊運用豐富的文字、影視語言和全方位的多媒體技術(shù),實現(xiàn)了圖文并茂、視頻與文本文件相結(jié)合。關(guān)于介入放射方面的專家論壇、基礎(chǔ)與臨床研究、短篇論著、個案和綜述等,以及手術(shù)錄像、講課幻燈均可以投稿。
現(xiàn)我刊特舉辦免費贈刊活動,贈刊對象可以是科室、圖書館、個人等,免費索取!
請從本刊網(wǎng)站http://zhjrfsxdzzz.paperopen.com“下載中心”處下載并填寫“贈刊索取單”,填好后發(fā)至郵箱zhjrfsx@163.com即可,我刊收到后會及時回復(fù)!
郵編:300190 地址:天津市南開區(qū)金平路10號2樓204室
電話:022-87087067 郵箱:zhjrfsx@163.com
網(wǎng)址:www.zhjrfsxdzzz.paperopen.com